Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
newtom_giano_hr_site_prep [2020/04/06 15:08]
joe.albertson
newtom_giano_hr_site_prep [2020/04/06 15:14] (current)
joe.albertson
Line 2: Line 2:
 ---- ----
 <​code>​PHYSICAL ROOM REQUIREMENTS </​code>​ <​code>​PHYSICAL ROOM REQUIREMENTS </​code>​
-The NewTom Giano HR requires a minimum physical space of x" wide, x" deep, and x" tall. The requirements for width and depth take into account the space needed for the patient to walk around the device and for the operator to use the device.+The NewTom Giano HR requires a minimum physical space of 60" wide, 63" deep, and 97" tall. The requirements for width and depth take into account the space needed for the patient to walk around the device and for the operator to use the device.
 <​bootnote warning>​24"​clearance on right side of machine is compulsory. Failure to meet this requirement will result in ergonomic and operational issues for clinic staff.</​bootnote>​ <​bootnote warning>​24"​clearance on right side of machine is compulsory. Failure to meet this requirement will result in ergonomic and operational issues for clinic staff.</​bootnote>​